Philips Addresses Patient Worn Monitor Holes

Wednesday, September 13, 2017 @ 08:09 AM gHale


Philips produced a software update that fixes one of two vulnerabilities in its IntelliView MX40 Patient Worn Monitor and provides mitigations for the other, according to a report with ICS-CERT.

Philips plans to release an additional software update in 2017 to address the remaining vulnerability.

RELATED STORIES
mySCADA Fixes myPRO Hole
Fix for Infusion Pump Issues in Jan.
Diabetes Management Software Hole Filled
SpiderControl Updates SCADA Web Server

IntelliVue MX40 Patient Worn Monitor (wireless local area networks WLAN only), all versions prior to Version B.06.18 suffer from the vulnerabilities.

Vulnerabilities only affect MX40 WLAN monitors operating on a user-provided 802.11 wireless LAN. MX40 monitors with 1.4 GHz and 2.4 GHz Smart-Hopping radios are not affected.

Successful exploitation of these vulnerabilities may allow an attacker to issue 802.11 Wi-Fi management commands that can impact reporting availability of MX40 device local monitoring to a central monitoring station. If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming (local audio-on, local screen-on) when it is not available from the local device, a delay of treatment can occur.

Philips is a global company that maintains offices in several countries around the world, including Africa, Asia, Europe, Latin America, the Middle East, and North America.

The MX40 patient worn monitor is intended to be used primarily as a traditional telemetry medical device as part of a monitoring and alerting system. As such, the MX40 (WLAN) connects to a central monitoring station by way of a wireless network to operate in compliance with required behaviors that ensure the central station maintains device status, and the local MX40 device continues to operate in local monitoring and alarming mode in the event of a Wi-Fi network access anomaly. When the MX40 is connected to a central monitoring station, the central monitor provides the primary patient monitoring and alarming function, and the local MX40 device operates in telemetry mode (local audio off, local screen off).

The MX40 is designed to:
• Operate in telemetry mode (local audio-off, local screen-off) to communicate status and alerts to hospital staff at the central monitoring station.
• Switch from telemetry mode into local monitoring mode (local audio-on, local screen-on) if certain Wi-Fi network management conditions or commands are encountered to de-authenticate MX40 from the Wi-Fi until either Wi-Fi network conditions resume to enable MX40 to reconnect and re-associate to central monitoring, or until hospital staff intervene to reset and reconnect the device.
• “Blacklist” the Wi-Fi access point in response to certain conditions as a security defense from further use until the device is reset by hospital staff.

The MX40 sees action across the healthcare and public health sector. Philips estimates these products are used primarily in the United States with users also in Canada, Europe, Australia, Africa, Latin America, Asia, and the Middle East.

Under specific 802.11 network conditions, a partial re-association of the MX40 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur.

CVE-2017-9657 is the case number assigned to this vulnerability, which has a CVSS v3 base score of 6.5.

In addition, certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the MX40 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff.

CVE-2017-9658 is the case number assigned to this vulnerability, which has a CVSS v3 base score of 6.5.

These vulnerabilities could be exploited via an adjacent network. No known public exploits specifically target these vulnerabilities. However, an attacker with a low skill would be able to exploit these vulnerabilities.

Philips released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

For additional information about the vulnerabilities or the proposed mitigations, contact Philips.

Philips is planning to release an additional MX40 software update in 2017 to address the improper handling of exceptional conditions vulnerability.

Please see the Philips product security web site for latest information for this and other Philips products.